Mean left ventricular diastolic, and systolic volumes indexed to body surface area and ejection fraction were: 24.7 ± 3.6 ml/m 2, 9.2 ± 1.3 ml/m 2 and 62 ± 3.8% using TomTec and 26.6 ± 3.8 ml/m 2, 10.4 ± 2 ml/m 2 and 63 ± 3.1% using QLAB, respectively. Left ventricular dimensions, volumes and ejection fraction were measured using Philips QLAB version 9.0 and results were compared to results of the same datasets, analysed using TomTec 4D LV software.

Philips qlab manual full#

38 babies underwent full Echocardiographic evaluation, including the acquisition of several 3-dimensional datasets, while naturally sleeping. Fifty consecutive newborns or neonates coming to their first newborn visit within the first 3 weeks of live, were recruited. The aim of this study was to compare left ventricular dimensions and ejection fraction obtained from the same 3-dimensional echocardiography datasets using 2 commercially available applications: Philips QLAB and Tomtec 4D LEFT VENTRICLE-ANALYSIS. The validity and normal values have been established in a large group of normal adults, children and neonates. Q-Station 3.0 with a2DQ and/or aCMQ Q-Apps are affected.Three-dimensional echocardiography is being used to evaluate left ventricular dimensions and ejection fraction in clinical practice. Intended to manage, view, analyze, and report qualitative and quantitative image data from ultrasound exams.

Philips qlab manual manual#

When using the QLAB Auto 2D Quantification (a2DQ) and Auto Cardiac Motion Quantification (aCMQ) applications, the reported End-Systolic Volume (ESV) may be smaller, and the Left Ventricular Ejection Fraction (EF) may be higher, than the ESV and EF calculated by manual tracing without the use of QLAB. An incorrect EF calculation could lead to misdiagnosis and/or delayed or incorrect therapy. Philips has identified additional affected units and sent the Urgent-Field Safety Notice -Philips Ultrasound QLAB 10.0 (on Xcelera) letter, dated 7, to their consignees. The FCO associated with this letter is FSN MA-FCO 83000190. The affected units identified as QLAB versions 10.0 with a2DQ and/or aCMQ plug-ins installed with Philips Xcelera software.

Philips qlab manual software#

QLAB software installed on Philips diagnostic ultrasound systems (carts) is the subject of a separate Field Safety Notice (FSN MA-FCO79500316 ).

Philips qlab manual upgrade#

Philips has corrected this issue with QLAB version 10.2 on Xcelera and will contact customers to arrange installation of the upgrade to version 10.2 or higher free of charge. If consignees need any further information or support concerning the FCO 83000190, please contact your local Philips representative or Philips Customer Service at 1-80.

philips qlab manual

Philips has begun sending out the URGENT - Medical Device Correction letters, dated 2014 NOV 05, for the Philips Ultrasound QLAB, Philips Model EPIQ 5 Ultrasound System, Philips Model EPIQ 7 Ultrasound System, and Philips Ultrasound Q-Station. The letters notified customers that Erroneous End-Systolic Volumes and Ejection Fraction May Be Reported When Using a2DQ & aCMQ (Auto 2D Quantification & Auto Cardiac Motion Quantification).Ĭustomers with questions should contact local Philips representative or Philips Customer Service at 1-86. Worldwide distribution (for FCO 83000190): US, including the states of AK, CA, CO, CT, DC, DE, FL, GA, ID, IL, IN, KS, KY, MA, MD, ME, MI, MN, MO, NC, NJ, NY, OH, OR, PA, SC, SD, TN, TX, VA, WA, and WI and the countries of United Arab Emirates, Austria, Australia, Belgium, Canada, Switzerland, Colombia, Germany, Ecuador, Estonia, Spain, Finland, France, Great Britain, Hong Kong, Israel, Italy, Japan, Luxembourg, Malaysia, Netherlands, Norway, New Zealand, Philippines, Saudi Arabia, Sweden, Singapore, Slovenia, Thailand, Turkey, Taiwan, and Vietnam.